EMF Study
(Database last updated on Mar 27, 2024)
ID Number |
|
738 |
Study Type |
|
Engineering & Physics |
Model |
|
2450 MHz (WiFi) transmitter exposure to medical devices and analysis of electromagnetic interference |
Details |
|
Medical Devices were tested using 2450 MHz LANaccess point transmitter operating at 500 milliwatts using the ANSI C63.18 ad hoc test protocol. Some EMI was observed in a physiological monitor, an ultrasound device, and an infusion pump. The authors suggest appropriate EMC characterization, management, and minimal seperation distances when employing hospilat 2.4 GHz LAN systems. |
Findings |
|
Not Applicable to Bioeffects |
Status |
|
Completed With Publication |
Principal Investigator |
|
St. Joseph's Hospital, WI, USA
|
Funding Agency |
|
Private/Instit.
|
Country |
|
UNITED STATES |
References |
|
Rice, WP J. Clin Engineering, (2000) Sep/Oct:260-264
|
Comments |
|
|
Return
|