EMF Study
(Database last updated on Mar 27, 2024)
ID Number |
|
214 |
Study Type |
|
Engineering & Physics |
Model |
|
30 kHz - 2.5 GHz (CW, GSM) exposure to cardiac pacemakers and analysis of electromagnetic interference |
Details |
|
Pacemakers (95 different types) were examined for EMI effects to assess potential risk for pacemakers wearers. Development of appropriate testing conditions allowed appropriate coupling of RF fields with the pacemaker. The majority of pacemakers operated without any significant EMI effects, although in some cases EMI effects were observed at certain frequencies. The authors recommend pacemaker patients not carry mobile phones that are switched on in their breast pocket, and during operation the unit should be held at a distance of 20cm or greater from the pacemaker implant site. There is no risk when the phone is placed to the ear at a distance of > 20cm from the pacemaker site.
|
Findings |
|
Not Applicable to Bioeffects |
Status |
|
Completed Without Publication |
Principal Investigator |
|
University of Gieben, FRG
|
Funding Agency |
|
FGF, Germany
|
Country |
|
GERMANY |
References |
|
|
Comments |
|
FGF database ID# D2, FGF contract# 5928 |